電子萬能試驗機操作規(guī)程
發(fā)布時間:2018-11-23 來源:濟南恒思盛大儀器有限公司 瀏覽:次
操作使用電子萬能試驗機有一定環(huán)境要求:保持測試環(huán)境整潔、干燥,附帶各夾具抗氧化性能良好;試驗機正常工作的環(huán)境溫度是10~35℃,相對濕度不大于80%。 在此環(huán)境下具體操作機器時有一定規(guī)程,我們簡單分享一下。
Electronic universal testing machine is operating environment requirements: to keep theenvironment clean, dry test, with antioxidant properties of each fixture is good; 10 ~ is a normal test machine environment temperature 35 ℃, relative humidity of not more than 80%.Under the environment of this specific operation machine has certain rules, we simply share.
首先,準(zhǔn)備工作
First of all, to work
1.試驗機進行可靠接地,地線接在試驗機的后部螺栓上。
1 test machine is reliable grounding, grounding on the rear bolt in the testing machine.
2.檢查試驗傳感器是否滿足試驗要求,是否需要更換傳感器,避免被試品測試力過大損壞傳
If the 2 test sensor to meet the test requirements, whether you need to replace the sensor,avoid sample measuring force is too big damage.
感器;或是因為傳感器超出被測試品數(shù)值過大超出誤差允許范圍。
Sensor; or because the sensor beyond the tested product value is too large to exceed the allowable error range.
然后,正式操作
Then, the formal operation
1. 按順序打開試驗機----電腦----打印機的電源開關(guān)(預(yù)熱15分鐘)。
1 in order to open the test machine, the power switch computer, printer (15 minutes).
2.待測試品放置試驗臺面,調(diào)節(jié)試驗機升降按鈕確保與測試面對準(zhǔn),并對試品做好裝夾。
The 2 tested product placement test table, adjust the lifting test machine and test button to ensure face alignment, and the sample do clamping.
3. 電腦打開軟件SANS-POWER TEST進入主界面;點擊試驗部分,進入編輯試驗方案界面。
3 open the computer software SANS-POWER TEST to enter the main interface; click the testpart, go to edit test program interface.
4.編輯試驗方案時,設(shè)置各試驗參數(shù)例如:測試項目,測試力值,限位值等;對軟件中原有試驗方案可直接使用,對新試驗方案需重新進行編輯和添加;保存退出。
The 4 edit test plan, set the test parameters such as: test project, test force value, limitingvalue; a test scheme can be used directly on the Central Plains of the new software, testprograms need to edit and add; exit last saved.
5.返回主界面后將“實時力值”,“位移”,“變形”,“峰值力值”進行清零。
5 to return to the main interface will "real-time force value", "displacement", "deformation","peak value" reset.
6.點擊右向三角按鈕開始試驗。
6 click on the right button to start the test to the river.
7.當(dāng)“實時力值”開始下降且“實時力值”明顯低于“峰值力值”時,可以點擊方框按鈕停止試驗,或利用試驗機自動返車(此項功能需在試驗方案中進行參數(shù)設(shè)置),記錄“峰值力值”比對標(biāo)準(zhǔn)做出試驗結(jié)果判定。
7 when the "real value" began to decline and "real-time force value" is obviously lower than that of "peak value", click on the box button to stop the test, or the testing machine automaticcar returning (this function need to set parameters in the test scheme), recording "peak force value" than the standard to make the test results judge.
8.斥卸工裝取下測試品,電腦退出SANS-POWER軟件并關(guān)機。
8 develop unloading tooling remove test products, computer software and shutdown exit SANS-POWER.
http://m.haratihotel.com/